Lakeshore Learning Materials 642 Manual de usuario Pagina 81

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 122
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 80
Lake Shore Model 642 Electromagnet Power Supply User’s Manual
Computer Interface Operation 5-29
Table 5-9. Command Summary
Command Function Page Command Function ......................................................Page
*CLS Clear Interface Cmd ..................................29 INTWR? Internal Water Mode Query ........................33
*ESE Event Status Enable Cmd..........................29 KEYST? Keypad Status Query ..................................33
*ESE? Event Status Enable Query........................30 LIMIT Limit Output Settings Cmd.........................34
*ESR? Event Status Register Query .....................30 LIMIT? Limit Output Settings Query.......................33
*IDN? Identification Query .................................. 30 LOCK Keypad Lock Cmd ......................................34
*OPC Operation Complete Cmd .........................30 LOCK? Keypad Lock Query....................................34
*OPC? Operation Complete Query .......................30 MAGWTR Magnet Water Mode Command..................34
*RST Reset Instrument Cmd...............................30 MAGWTR? Magnet Water Mode Query ........................34
*SRE Service Request Enable Cmd ....................31 MODE IEEE Interface Mode Cmd..........................34
*SRE? Service Request Enable Query..................31 MODE? IEEE Interface Mode Query .......................34
*STB? Status Byte Query .....................................31 OPST? Operational Status Query............................35
*TST? Self-Test Query.........................................31 OPSTE Operational Status Enable Cmd ..................35
*WAI Wait-To-Continue Cmd ............................31 OPSTE? Operational Status Enable Query................35
BAUD RS-232C Baud Rate Cmd .........................31 OPSTR? Operational Status Register Query..............35
BAUD? RS-232C Baud Rate Query .......................31 RATE Current Ramp Rate Setting Cmd ................35
DFLT Factory Defaults Cmd ...............................32 RATE? Current Ramp Rate Setting Query ..............35
DISP Display Parameter Cmd ............................32 RDGI? Current Output Reading Query ...................35
DISP? Display Parameter Query ..........................32 RDGV? Output Voltage Reading Query...................36
ERCL Error Clear Cmd........................................32 RSEG Ramp Segments Enable Cmd......................36
ERST? Error Status Query .................................... 32 RSEG? Ramp Segments Enable Query ...................36
ERSTE Error Status Enable Cmd...........................32 RSEGS Ramp Segments Parameters Cmd ...............36
ERSTE? Error Status Enable Query ........................32 RSEGS? Ramp Segments Parameters Query.............36
ERSTR? Error Status Register Query ......................33 SETI Output Current Setting Cmd .......................36
IEEE IEEE-488 Interface Parameter Cmd..........33 SETI? Output Current Setting Query .....................36
IEEE? IEEE-488 Interface Parameter Query........33 STOP Stop Output Current Ramp Cmd.................37
INTWR Internal Water Mode Command ..............33 XPGM External Program Mode Cmd .....................37
XPGM? External Program Mode Query ...................37
5.3.1 Interface Commands (Alphabetical Listing)
*CLS Clear Interface Command
Input: *CLS[term]
Remarks: Clears the bits in the Status Byte Register and Standard Event Status Register and terminates all
pending operations. Clears the interface, but not the instrument. The related instrument command
is *RST.
*ESE Standard Event Status Enable Register Command
Input: *ESE <bit weighting>[term]
Format: nnn
Remarks: The Standard Event Status Enable Register determines which bits in the Standard Event Status
Register will set the summary bit in the Status Byte. This command programs the enable register using
a decimal value that corresponds to the binary-weighted sum of all bits in the register. Refer to
Paragraph 5.1.4.2.1.
Vista de pagina 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 121 122

Comentarios a estos manuales

Sin comentarios