4.0. GENERAL
Either of two computer interfaces on the optional 6224 card permit remote MPS operation: the IEEE-488
Interface (Paragraph 4.1) and the Serial Interface (Paragraph 4.2). Use only one of the interfaces at a time.
The two interfaces share a common set of commands described in Paragraph 4.3. For further information on
RS-485 Mainframe Control Bus Operation, see Appendix C.
4.1. IEEE-488 INTERFACE
The IEEE-488 Interface is an instrumentation bus with hardware and programming standards that simplify
instrument interfacing. The MPS IEEE-488 Interface complies with the IEEE-488.2-1987 standard and
incorporates the functional, electrical, and mechanical specifications of the standard unless otherwise
specified in this manual.
All instruments on the interface bus perform one or more of the interface functions of TALKER, LISTENER, or
BUS CONTROLLER. A TALKER transmits data onto the bus to other devices. A LISTENER receives data
from other devices through the bus. The BUS CONTROLLER designates to the devices on the bus which
function to perform. The MPS performs the functions of TALKER and LISTENER but cannot be a BUS
CONTROLLER. The BUS CONTROLLER is the digital computer which tells the MPS which functions to
perform.
Below are MPS IEEE-488 interface capabilities:
SH1: Source handshake capability SR1: Service request capability
RL1: Complete remote/local capability AH1: Acceptor handshake capability
DC1: Full device clear capability PP0: No parallel poll capability
DT0: No device trigger capability
C0: No system controller capability
T5: Basic TALKER, serial poll capability, talk only, unaddressed to talk if addressed to listen
L4: Basic LISTENER, unaddressed to listen if addressed to talk
4.1.1 IEEE-488 Interface Settings
To use the IEEE-488 interface, the user must set
the IEEE Address and Terminators. Press the
Function Menu key, then press the key
corresponding to INTERFACE SETUP to display the
Interface Setup screen. Use the cursor keys to
move the arrow indicator to IEEE ADD and
TRMS/EOI.
IEEE ADD: 1 to 30
Use the Data Entry up or down arrows to increment or decrement the IEEE Address. The default value is 12.
The range of choices is from 1 to 30.
TRMS/EOI: CR LF, LF CR, DAB, LF, with or without EOI
Use the Data Entry up or down arrows to cycle through the following choices: (CR LF) EOI, (LF CR) EOI,
(DAB) EOI, (LF) EOI, (CR LF), (LF CR), (DAB), (LF). If turned on, End Or Identify (EOI) is asserted during the
last byte of a multibyte transfer. The default is Carriage Return and Line Feed (CR LF) with EOI ON.
INTERFACE SETUP:
¤ IEEE ADD
TRMS/EOI
RS232C
CB DEF
CB ADD
VW ANGLE
MF ID
OUTPUTS: *
+1.999A
: 12
: (CR LF) EOI
: 9600,N,8,1
: 9600,0,7,1
: 0
: 9
: 622 (1)
¥
¥
¥
¥
¥